Battery Case 2. RS232/485 Input/Output 3. DC 9V Power Input 4. Calibration Switch 5. Load Cell Connect Socket FM User Manual 8 SME300000119 EXCELL PRECISION CO., LTD. Installation 4-1 Load Cell 12345 6789 Load cell cable Shield Load cell EXC+ SEN+ SIG+ SIGEXCSEN- 4-wired (5-wired) Load Cell Short Pin4&5 to connect with EXC+ Short Pin2&3 to connect with EXC- Pin1 connects with SIG+ Pin9 connects with SIG- Pin6, 7, 8 connect with Shield 6-wired (7-wired) Load Cell Pin5 connects with EXC+ Pin4 connects with SEN+ Pin3 connects with EXC- Pin2 connects with SEN- Pin1 connects with SIG+ Pin9 connects with SIG- Pin6, 7, 8 connect with Shield FM User Manual 9 SME300000119 EXCELL PRECISION CO., LTD. FM User Manual 10 SME300000119 4-2 Dimension EXCELL PRECISION CO., LTD. FM User Manual 11 SME300000119 Stu.ff .th.e.su.rp.lu.s.wi.re.s i.nt.o the battery case. 4-3 Battery Assembly EXCELL PRECISION CO., LTD. Internal Calibration Adjust calibration switch to “ON”, and the screen displays: After 2 seconds Press Press Press Press Adjust calibration switch back to “OFF”. ZERO TARE ZERO TARE ENTER ENTER ENTER . Specification Setting . Internal Weight Calibration . Internal Function Setting FM User Manual 12 SME300000119 EXCELL PRECISION CO., LTD. 5-1 Specification Setting Press ENTER ENTER Input the *Parameter Code intended Press *Parameter Code . Decimal Point . Capacity . Division 1 . Division 2 . Zero Tracking Setting . Unstable Detecting Setting The screen displays the parameters set previously. ENTER Input the parameters intended and then press Continue other functions setting or press E SC to escape. ZERO TARE GROSS NET 3 F2 4 ENTER ESC . To add the value flashing . To reduce the value flashing . To move the cursor leftward . To move the cursor rightward . Storage setting . To abort setting / escape FM User Manual 13 SME300000119 EXCELL PRECISION CO., LTD. • Specification Parameter Description Parameter Code Function Setting Default Setting Parameter Description CSP-00 Decimal Point Refer to the description on next page. CSP-01 Capacity 999999 . 000000 Max. value for weight display 999999 CSP-02 Division 1 1 Min. value for weight display 1 2 5 10 20 50 CSP-03 Division 2 1 Min. value for Weight display 1 2 5 10 20 50 CSP-04 Zero Tracking Setting Refer to the description on next page. CSP-05 Unstable Detecting Setting Refer to the description on next page. FM User Manual 14 SME300000119 EXCELL PRECISION CO., LTD. • Parameter Display Description CSP-00 Decimal Point Display Decimal Point Digit None . 1 Digit . 2 Digits . 3 Digits . 4 Digits CSP-04 Zero Tracking Setting Display Division/Period . 0.25 D/1 sec . 0.5 D/1 sec . 0.75 D/1 sec 1D/1 sec . 1.25 D/2 sec . 1.5 D/2 sec . 1.75 D/2 sec 2 D/2 sec No Zero Tracking CSP-05 Unstable Detecting Setting Display Division / Period . 0.25 D/1 sec . 0.5 D/1 sec . 0.75 D/1 sec 1D/1 sec . 1.25 D/2 sec . 1.5 D/2 sec . 1.75 D/2 sec 2 D/2 sec No Unstable Detecting FM User Manual 15 SME300000119 EXCELL PRECISION CO., LTD. 5-2 Internal Weight Calibration Turn on and warm up the machine for 15 to 30 minutes before calibration. Adjust calibration switch to “ON”, and the screen will display . Press TARE or ZERO to select . Procedure Zero Calibration ENTER ENTER Zero calibration complete Weight Calibration Confirmed Place the weight mass and set the weight ENTER ENTER ESC ESC Calibration Switch “OFF” Zero Calibration a ) Ensure nothing on the platter; after being stable, press ENTER , and the screen will display “ ……. ”. The zero calibration will be complete 5 seconds later. b ) To abort zero calibration, just press . ESC Weight Calibration a ) Place an object, whose weight is known, on the platter, and input the weight value from ENTER ESC front panel. After being stable, press and the screen will display “……. ”. The weight calibration will be complete 5 seconds later. b ) To abort weight calibration, just press . FM User Manual 16 SME300000119 EXCELL PRECISION CO., LTD. • Password Setting Adjust calibration switch to “ON” After 2 seconds ENTER ENTER F1 ZERO Input new password • “0000” means no password. Confirm new password (re-input again) • After complete password setting, when entering calibration mode or function setting mode, the screen displays for 1 second, and then . It’s necessary to input the correct password to continue each setting. If the password inputted is not correct, the screen displays . FM User Manual 17 SME300000119 EXCELL PRECISION CO., LTD. 5-3 Internal Function Setting Press Input the *Parameter code intended ENTER *Parameter Code . Tare or Zero Function under Unstable Status . Turning on Re-zero . Re-zero Range . Filter Strength Press ENTER . AD Sampling Rate . Animal Scale Mode The screen displays . Animal Scale Stable Range the parameters set previously. Input the parameters intended ENTER . Animal Scale Sampling and then press Frequency . Dual Range Resolution Setting . Dual Range Resolution Middle Point Setting Continue other functions sett...